CANNON-MANNING SEMI-MICRO EXTRA LOW CHARGE VISCOMETER
CANNON CM Semi-Micro ELC GV

The Cannon-Manning Semi-Micro Extra Low Charge Viscometer measures kinematic viscosity of very small samples of transpar...

 

CANNON-Manning Semi-Micro Extra Low Charge Viscometer

CANNON-Manning Semi-Micro Extra Low Viscometers

Catalog Number Size Approximate Constant
(mm²/s²)
Viscosity Range (mm²/s)†
Uncalibrated Calibrated Low High
9722-C50 9722-D50 25 0.002 0.4 2.0
9722-C53 9722-D53 50 0.004 0.8 4.0
9722-C56 9722-D56 75 0.008 1.6 8.0
9722-C59 9722-D59 100 0.015 3.0 15
9722-C62 9722-D62 150 0.035 7.0 35
9722-C65 9722-D65 200 0.1 20 100
9722-C68 9722-D68 300 0.25 50 250
9722-C71 9722-D71 350 0.5 100 500
9722-C74 9722-D74 400 1.2 240 1200
9722-C77 9722-D77 450 2.5 500 2500
9722-C80 9722-D80 500 8.0 1600 8000
9722-C83 9722-D83 600 20 4000 20000

† Viscosity ranges in mm²/s represent flow times of 200 to 1000 seconds

The Cannon-Manning Semi-Micro Extra Low Charge Viscometer measures kinematic viscosity of very small samples of transparent Newtonian liquids according to ASTM D445 and ISO 3104. Minimum sample volume is ~0.5 mL (sizes 200 through 600 may require additional sample). Minimum bath depth is 230 mm (9 in). Both calibrated and uncalibrated viscometers are available. Calibrated viscometers include a certificate of calibration and instruction sheet. Holders are not available.